Customization: | Available |
---|---|
After-sales Service: | 1 Year |
Warranty: | 1 Year |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
System performance parameters AutoPure Pilot600 m performance parameter
Main system size | L*W*H 600x500x850mm | |
Main system weight | 65kg | |
Control system | Unique CDSystem | |
Enclosure rating | IP54 | |
Electricity supply | 100~ 240 V, 50 ~ 60 Hz | |
Working environment | Temperature: 4ºC~35ºC, Relative Humidity: 20%~95%, no condensation |
|
System pump (Dual pump system) |
A/B pump Flow rate range: 4~600ml/min (2- pump packing mode: 1~ 1200ml/min) Eligibility range: 5%~95% Pressure range: 0~ 2MPa (20bar, 290 psi) Viscosity: 0.7~10 cp |
|
UV detector |
280nm fixed wavelength detector LED light source, lifespan>8000h Detection range: - 6~ +6AU Linearity: between±2%@0 ~ 2AU Wavelength Accuracy: ±1nm Wavelength Repeatability: ±0.1nm Noise:±2*10 - 5AU |
DAD full spectrum direct reading detector, 200~400nm variable dual wavelength. Wavelength Accuracy: ±1nm Wavelength Repeatability: ±1 nm Noise: ±4*10 - 5AU (@254nm,1S) |
Optical fiber conduction, external flow cell, avoid light source temperature affecting the activity of protein samples, double beam design with high stability. Standard flow cell: 2mm optical diameter, 50μl flow cell volume, maximum pressure resistance 2Mpa |
||
Conductivity detector | Detection range: 0.1mS/cm ~ 999.9mS/cm Detection cell volume: 20μl Accuracy: ±3% or 0.2mS/cm built-in temperature sensor, automatic compensation calibration |
|
Temperature detector | Temperature range: 0 ~99ºC; Temperature Accuracy:± 2ºC 2ºC ~50ºC |
|
pH detector | Detection range:2~ 12;Accuracy: ±0.15 | |
Inlet options | 4 inlets for single pump A total of 8 inlets of double pumps Expandable to 12 inlets |
|
Outlets options | Maximum 8 outlets | |
Column valve | Single column | |
Bubble trap | Effectively remove air bubbles in flowpath, optional online or offline | |
Mixer | Mixing chamber volume: 12ml | |
Bubble sensor | Support up to 3 | |
Pressure sensor | One after each pump(A,B), one before the column, one after the column, a total of four, detection range is 0- 2MPa |